PL-ELS 2100 Evaporative Light Scattering Detector (ELSD)

移動相揮發光散式檢測器

The PL-ELS 2100 surpasses other ELSDs for low temperature HPLC applications with semi-volatile compounds.

The PL-ELS 2100 represents the next generation of ELSD technology. Its innovative design provides optimum performance across a diverse range of HPLC applications. This patented design permits operation at very low temperatures, even ambient, to facilitate the detection of semi-volatile compounds not seen by other ELSDs.


In ELSDs, the eluent is removed as it passes through the evaporator tube. The temperature of this tube is critical in optimizing the detection of small molecules. If the temperature is too high, thermally labile compounds in the sample may evaporate along with the eluent and will not be detected. The PL-ELS 2100 can evaporate high boiling solvents at very low temperatures, for example, a mobile phase consisting of 100% water at 30°C. The novel design (patent pending) of the PL-ELS 2100 provides superior performance compared to competitors' detectors for the analysis of semi-volatile compounds.

 
 

 

PL-ELS 2100 Applications

Analysis of Parabens
This application shows a separation of parabens, synthetic preservatives frequently used in cosmetics and personal care products and in pharmaceutical formulations. These low molecular weight compounds are relatively volatile but can easily be detected by operating the PL-ELS 2100 at very low temperatures, in this case, 30°C.

Column: Adsorbosil C18 5µm, 150x4.6mm Key
1. Methyl paraben
2. Ethyl paraben
3. Propyl paraben
Eluent A: Water
Eluent B: ACN
Gradient: 50-75% B in 7 mins
Flow Rate: 1.0ml/min
Inj Vol: 10µl
Sample Conc: 1mg/ml
Detector: PL-ELS 2100 (neb=30°C, evap=30°C, gas=1.0 SLM)

Analysis of Parabens (ELSD)
 


Column: Adsorbsil C18 5µm, 150x4.6mm Key
1. Acetanilide
2. Indapamide
3. Ibuprofen
4. Dibutylphthalate
Eluent A: 0.1% TFA in Water
Eluent B: 0.1% TFA in ACN
Gradient: 60-90% B in 5 mins
Flow Rate: 1.0ml/min
Inj Vol: 20µl
Sample Conc: 1mg/ml
Detector: PL-ELS 2100 (gas=1.6 SLM)

Analysis of Pharmaceutical Compounds (ELSD)



Analysis of Non-Volatile Peptides

Column: PLRP-S 100Å 5µm, 150x4.6mm Key
1. Oxytocin
2. Angiotensin II
3. Angiotensin I
4. Insulin
Eluent A: 0.1% TFA in Water
Eluent B: 0.1% TFA in ACN
Gradient: 20-50% B in 15 mins
Flow Rate: 1.0ml/min
Detector: PL-ELS 2100 (neb=90°C, evap=90°C, gas=1.8 SLM)


Analysis of Non-Volatile Peptides
 


Column: PLRP-S 100Å 5µm, 50x4.6mm Key
1. Indapamide
2. Dibutylphthalate
Eluent A: 0.05% TFA in Water
Eluent B: 0.05% TFA in ACN
Gradient: 5-95% B in 1 min
Flow Rate: 5.0ml/min
Inj Vol: 10µl
Sample Conc: 0.5mg/ml
Detector: PL-ELS 2100 (neb=30°C, evap=30°C, gas=1.6 SLM)

High Throughput Screening (ELSD)


Column: Xterra C18 3.5µm, 30x2.1mm Key
1. Dipropylphthalate
2. Dibutylphthalate
3. Dipentylphthalate
4. Diheptylphthalate
5. Dioctylphthalate
Eluent A: Water
Eluent B: ACN
Gradient: 0-100% B in 3.1 mins, hold for 0.7 min
Flow Rate: 0.5ml/min
Inj Vol: 10µl
Sample Conc: 0.1mg/ml
Detector: PL-ELS 2100 (neb=25°C, evap=25°C, gas=1.6 SLM)

Analysis of Phthalates (ELSD)
 

 

 